TESTING GUIDE

Separate purity from quantity

Research and quality-control information.

A ratio is not a mass

Chromatographic purity usually expresses a target peak’s share of integrated detector response under a stated method. It is not necessarily a mass fraction of every substance in the vial.

Content needs calibration

A quantitative assay uses an appropriate calibration and preparation basis to estimate the amount of the target analyte. The report must state units and the calculation basis.

Read both together

A strong purity result can coexist with a low amount. Neither result alone establishes sterility, safety, or suitability for a particular use.